Go Back Advanced Search Page

Analysis of Tagatose According to USP Method (SC1011)

According to USP 40 method, the assay of tagatose should be used a column packed with L22. It is necessary for the system suitability to satisfy relative standard deviation (RSD) of ≤ 2.0%. It was confirmed to satisfy the system suitability when tagatose was analyzed using SUGAR SC1011.


Sample : 20µL
1. Tagatose 5mg/mL

 

Column       : Shodex SUGAR SC1011 (8.0mmI.D. x 300mm)
Eluent       : 0.05mg/mL Calcium acetate aq.
Flow rate    : 0.6mL/min
Detector     : Shodex RI
Column temp. : 85°C