KF-600 | · Approximately half of the analysis time compared with standard columns · The amount of solvent used is reduced to about a third · Improved applicability of solvent replacement · Corresponds to USP L21 |
---|
Please refer to Replacement of In-Column Solvent
*The columns with ‘M’ at the end of column names are mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution.
*Use of the KF-600 series with semi-micro type devices is recommended.
(Rapid analysis downsized columns)
KF-600 series shipping solvent : Tetrahydrofulan (THF)
Product code | Product name | Plate number (TPN/column) |
Particle size (μm) |
Size (mm) I.D. x Length |
---|---|---|---|---|
F6028091
F6028092 F6028093 F6028094 F6028095 F6028096 F6028097 F6028098 F6028099 F6700300 |
GPC KF-601 GPC KF-602 GPC KF-602.5 GPC KF-603 GPC KF-604 GPC KF-605 GPC KF-606 GPC KF-606M GPC KF-607 GPC KF-G 4A (GPC KF-G) |
≥ 17,000
≥ 17,000 ≥ 17,000 ≥ 17,000 ≥ 16,000 ≥ 7,000 ≥ 7,000 ≥ 8,000 ≥ 5,000 Guard column |
3 3 3 3 3 10 10 10 18 8 |
6.0 x 150 6.0 x 150 6.0 x 150 6.0 x 150 6.0 x 150 6.0 x 150 6.0 x 150 6.0 x 150 6.0 x 150 4.6 x 10 |
For all products
Max. usable pressure: (KF-601 to KF-604) 6.0MPa, (KF-605 to KF-607) 0.6MPa
Recommended flow rate: 0.4 to 0.6mL/min
Max. usable flow rate: 0.8mL/min
Max. usable temperature: 45°C
[Target molecular weight range and Exclusion limit]
* Please use the above table as a rough indication for the column selection.
• With Polystyrene (eluent : THF)
Product name | Target molecular weight range | Exclusion limit |
---|---|---|
KF-601 KF-602 KF-602.5 KF-603 KF-604 KF-605 KF-606 KF-606M KF-607 |
100 – 700 200 – 1,500 300 – 10,000 600 – 50,000 7,000 – 500,000 50,000 – 2,000,000 150,000 – *(20,000,000) 1,000 – *(20,000,000) 300,000 – *(200,000,000) |
1,500 |
*( ) Estimated value